GC Column Cooling Program Reduces Bleed Carryover

Resolve Bottlenecks,
Find Innovative Solutions
Generate Solutions

Solution Overview

Problem

Gas chromatography (GC) systems face challenges in reducing the effects of column bleed carryover, which occurs due to thermally induced decomposition of the stationary phase, leading to interference with sample analysis and reduced sample throughput.

Innovation Solution

Implementing a three-stage column cooling program that includes a high-speed cooling ramp, an isothermal dwell time at a controlled temperature, and a second high-speed cooling ramp to effectively flush and remove decomposition products, while optionally increasing carrier gas flow during the dwell time to enhance flushing efficiency.

Engineering Contradictions & Design Principles

VSEngineering Contradiction Analysis

1Productivity

If the column is cooled rapidly after a sample run, then the sample throughput is improved, but the column bleed carryover increases causing peak interference in the chromatogram

Engineering Contradiction:
Improvesample throughputVSAvoidcolumn bleed carryover
Core Design Contradiction:
ProductivityVSObject-generated harmful factors

Solution Approach 1:

The cooling process is segmented into multiple stages: a first high-speed cooling ramp from initial final temperature to an intermediate dwell temperature, followed by an isothermal dwell period, then a second high-speed cooling ramp to the initial temperature. This segmentation allows the column to be cooled efficiently while managing bleed carryover at each stage.

Inventive Principle:
Principle #1Segmentation

Solution Approach 2:

The isothermal dwell period at the intermediate temperature serves as a preliminary action before the final cooling step. During this dwell time, the carrier gas flow is increased to flush and remove decomposition products from the column before the temperature is lowered further, preventing these products from causing interference in subsequent runs.

Inventive Principle:
Principle #10Preliminary action

2Object-generated harmful factors

If the column is cooled slowly and evenly, then the column bleed carryover is reduced, but the sample throughput decreases

Engineering Contradiction:
Improvecolumn bleed carryoverVSAvoidsample throughput
Core Design Contradiction:
Object-generated harmful factorsVSProductivity

Solution Approach 1:

The cooling process is divided into segments where high-speed cooling is used for the majority of the temperature range, and a controlled dwell is inserted only at the intermediate temperature point. This allows most of the cooling to occur rapidly while still managing bleed carryover effectively.

Inventive Principle:
Principle #1Segmentation

Solution Approach 2:

The method rushes through the temperature ranges where column bleed is either fully retained or unretained by using high-speed cooling ramps. The dwell period is strategically placed only where it is most effective, skipping unnecessary slow cooling phases and thus maintaining high sample throughput.

Inventive Principle:
Principle #21Skipping (Rushing through)

3Object-generated harmful factors

If the carrier gas flow is increased during cooling, then the flushing of decomposition products is improved, but the energy consumption increases

Engineering Contradiction:
Improvedecomposition product removalVSAvoidcarrier gas consumption
Core Design Contradiction:
Object-generated harmful factorsVSUse of energy by moving object

Solution Approach 1:

The increased carrier gas flow is applied as a preliminary flushing action during the isothermal dwell period at the intermediate temperature, before the final cooling to initial temperature. This timing ensures that decomposition products are removed when they are most mobile, maximizing flushing efficiency while limiting the duration of high gas consumption.

Inventive Principle:
Principle #10Preliminary action

Solution Approach 2:

The carrier gas flow is increased only partially (during the dwell period) rather than continuously throughout the entire cooling process. This partial application of excessive flow achieves sufficient flushing of decomposition products while avoiding unnecessary energy consumption during other phases of the cooling cycle.

Inventive Principle:
Principle #16Partial or excessive action

Applied Scientific Principles

This section explains which scientific principles are used to turn an abstract innovation direction into a practical engineering solution.

Function Achieved in This Case

This approach significantly reduces baseline noise and peak interference, improving analysis accuracy and maintaining high sample throughput by effectively managing column bleed carryover.

Implementation Method 1

heating the GC column according to a heating program, the heating program comprising increasing a column temperature of the GC column from an initial column temperature to a final column temperature

Methodology Applied
Scientific EffectHeating: Heating

Implementation Method 2

cooling the GC column according to a cooling program, the cooling program comprising: decreasing the column temperature from the final column temperature to a dwell temperature

Methodology Applied
Scientific EffectCooling: Cooling

Implementation Method 3

flowing a carrier gas through a GC column; during a sample run time, heating the GC column... injecting the sample into the flowing carrier gas... and flowing the mixture through the GC column

Methodology Applied
Scientific EffectGas flow: Convection

Implementation Method 4

During column flow the sample encounters a stationary phase (typically a material lining the inside surface of the column), which causes different components of the sample to separate according to different affinities with the stationary phase

Methodology Applied
Scientific EffectChromatography: Chromatography

Implementation Method 5

as the temperature increases the stationary phase can begin to degrade. As it degrades, decomposition products are produced, and these decomposition products are volatile enough that at higher temperatures they are unretained or partially retained by the stationary phase

Methodology Applied
Scientific EffectThermal decomposition: Decomposition (biological)

Data Source

PatentUS11009489B2Method and system for reducing the effects of column bleed carryover
Publication Date: 2021.05.18 AGILENT TECHNOLOGIES INC
  • US11009489B2 patent drawing
  • US11009489B2 patent drawing
  • US11009489B2 patent drawing

AI summary

In gas chromatography (GC), a sample is introduced into a flow of carrier gas and the mixture is driven through a heated GC column to acquire chromatographic data from the sample. During this time, the column is heated from an initial temperature to a final temperature. Subsequently, the column is cooled according to a cooling program. The cooling program may include a first cooling ramp, a subsequent isothermal hold, and a subsequent second cooling ramp. Alternatively, while the column is cooled down the flow of carrier gas through the column may be slowed down or ceased for a period of time, after which the flow of carrier gas through the column may be resumed at the original flow rate in preparation for processing another sample. Controlling column temperature and/or flow in this manner may be effective for reducing column bleed carryover and/or the effects thereof.