Spectrophotometer Contamination Detection via Reference Spectrum Comparison

Resolve Bottlenecks,
Find Innovative Solutions
Generate Solutions

Solution Overview

Problem

Current spectrophotometric methods for determining hemoglobin derivatives in blood are prone to errors due to contaminants in the measuring cuvette, which can lead to incorrect concentration calculations, as the absorption of contaminants is not accurately accounted for, especially when the cuvette is not completely cleaned between sample and reference measurements.

Innovation Solution

A method that compares the reference spectrum of a measuring cuvette to a known target spectrum to detect contamination, using comparison parameters and predefined threshold values to determine if a cuvette is contaminated, allowing for automatic initiation of cleaning or correction measures without additional apparatus, and utilizing techniques like cross-correlation for pattern recognition to enhance sensitivity.

Engineering Contradictions & Design Principles

VSEngineering Contradiction Analysis

1Productivity

If the cuvette is not completely cleaned between sample and reference measurements, then measurement speed is improved, but measurement precision deteriorates due to contaminant absorption

Engineering Contradiction:
Improvemeasurement speedVSAvoidconcentration calculation accuracy
Core Design Contradiction:
ProductivityVSMeasurement precision

Solution Approach 1:

The system performs a preliminary reference measurement with clean cuvette to establish a target spectrum, then compares subsequent reference measurements against this baseline to detect contaminants before they affect sample measurements

Inventive Principle:
Principle #10Preliminary action

Solution Approach 2:

The system continuously monitors the reference spectrum against the target spectrum and provides feedback when contamination is detected, allowing for automatic correction or warning before inaccurate measurements are recorded

Inventive Principle:
Principle #23Feedback

2Measurement precision

If additional cleaning steps are added to detect and remove contaminants, then measurement precision is improved, but device complexity increases

Engineering Contradiction:
Improvereference spectrum accuracyVSAvoidcleaning process complexity
Core Design Contradiction:
Measurement precisionVSDevice complexity

Solution Approach 1:

The system uses the existing spectrophotometer hardware to perform self-diagnosis by measuring the reference spectrum and comparing it against the stored target spectrum, eliminating the need for external contamination detection devices

Inventive Principle:
Principle #25Self-service

Solution Approach 2:

The system creates a digital copy (target spectrum) of the clean cuvette reference measurement and uses this template to detect deviations caused by contaminants, avoiding the need for physical cleaning verification

Inventive Principle:
Principle #26Copying

3Measurement precision

If the reference spectrum is compared to a known target spectrum, then contaminant detection accuracy is improved, but measurement time increases

Engineering Contradiction:
Improvecontaminant detection accuracyVSAvoidmeasurement cycle time
Core Design Contradiction:
Measurement precisionVSLoss of time

Solution Approach 1:

The system performs spectral comparison only at critical wavelengths where contaminants typically absorb, rather than analyzing the entire spectrum, reducing computational time while maintaining detection accuracy

Inventive Principle:
Principle #16Partial or excessive action

Solution Approach 2:

The target spectrum is pre-stored in memory during initial calibration, eliminating the need for time-consuming reference measurements during each sample analysis cycle

Inventive Principle:
Principle #10Preliminary action

Applied Scientific Principles

This section explains which scientific principles are used to turn an abstract innovation direction into a practical engineering solution.

Function Achieved in This Case

This method enables accurate detection of contaminants, ensuring reliable concentration calculations by differentiating between true reference spectra and those affected by contaminants, thereby improving the reliability of spectrophotometric measurements and reducing errors in concentration values.

Implementation Method 1

The concentration calculation of the hemoglobin derivatives is based on the measured summation absorption of the individual components at multiple wavelengths: Aλ=log (I0λ/Iλ)=ΣεIλ*ci*d

Methodology Applied
Scientific EffectLambert-Beer law: Absorption (EM radiation)

Implementation Method 2

Spectrometers used for this purpose frequently operate according to the polychromator principle, i.e., in this method the incident light is first split using a polychromator into its spectral components

Methodology Applied
Scientific EffectPolychromator principle: Dispersion (of waves)

Implementation Method 3

The transmitted light is conducted to the polychromator by a further optical fiber, where it is split into its spectral components and imaged on the surface of a photosensitive receiver (CCD sensor)

Methodology Applied
Scientific EffectPhotoelectric effect: Photoelectric Effect

Data Source

PatentUS8379202B2Method for detecting contaminants
Publication Date: 2013.02.19 ROCHE DIAGNOSTICS OPERATIONS INC
  • US8379202B2 patent drawing
  • US8379202B2 patent drawing
  • US8379202B2 patent drawing

AI summary

A method for the detection of contaminants in an optical measuring cuvette of a spectrophotometer, typically an oximeter for determining hemoglobin derivatives, is provided, in which measuring cuvette, in addition to at least one sample measurement to obtain a sample spectrum I(λ), at least one reference measurement is performed using a reference liquid to obtain a reference spectrum I0(λ). The reference spectrum I0(λ) is compared to a known target spectrum I0soll(λ), which is associated with the measuring cuvette, comparison parameters being obtained, it being decided automatically whether a contamination of the measuring cuvette exists as a function of predefined threshold values of the comparison parameters.