Detecting method of ethene diamine and method for detecting ethene diamine in sitagliptin phosphate intermediate XG-SM2
A technology of XG-SM2 and detection method, which is applied in the field of analytical chemistry, can solve problems such as the detection of ethylenediamine that have not been seen before, and achieve the effect of good quality, high sensitivity and strong operability
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0085] Example 1: Selection of mobile phase
[0086] The purpose of this example is to investigate the influence of methanesulfonic acid as the mobile phase, its concentration and the elution procedure on the separation and detection of ethylenediamine.
[0087] Except for the mobile phase, the following chromatographic conditions were used in this example
[0088] Ion chromatograph: Thermo Dionex Aquion type ion chromatograph;
[0089] Chromatographic column: Thermo Dionex IonPac CS12A cation column (4mm*250mm);
[0090] Guard column: Thermo Dionex IonPac CG12 guard column (4mm*50mm);
[0091] Flow rate: 1.0ml / min;
[0092] Column temperature: 28-35℃;
[0093] Detector: Conductivity detector
[0094] Injection volume: 25μl.
[0095] In this embodiment, the ethylenediamine reference substance solution is used for determination. The preparation method of the ethylenediamine reference substance solution is as follows:
[0096] Accurately measure an appropriate amount of ethylenediamine, dissolv...
Embodiment 2
[0138] Example 2 Methodological study of the ethylenediamine detection method of the present invention
[0139] On the basis of Example 1, a detection method for ethylenediamine was established, specifically:
[0140] I. Chromatographic conditions:
[0141] Ion chromatograph: Thermo Dionex Aquion type ion chromatograph;
[0142] Chromatographic column: Thermo Dionex IonPac CS12A cation chromatographic column (4mm*250mm);
[0143] Guard column: Thermo Dionex IonPac CG12 guard column (4mm*50mm);
[0144] Mobile phase: 40mmol / L methanesulfonic acid (prepared with ultrapure water)
[0145] Flow rate: 0.7-1.4ml / min, preferably 1.0ml / min;
[0146] Column temperature: 28-35℃;
[0147] Detector: Conductivity detector
[0148] Injection volume: 25μl.
[0149] II. Preparation of reference solution:
[0150] Accurately measure an appropriate amount of ethylenediamine, dissolve it with ultrapure water and dilute it to a solution containing about 2μg ethylenediamine per 1ml, and get it;
[0151] Preparatio...
Embodiment 3
[0194] Example 3 Use the method of the present invention to determine two batches of sitagliptin phosphate intermediate XG-SM2
[0195] The specific measurement results are as follows:
[0196] Weigh appropriate amount of sitagliptin phosphate intermediate XG-SM2, dissolve and dilute with ultrapure water to make a solution containing about 1mg per 1ml, as a reference solution; accurately measure an appropriate amount of ethylenediamine and dilute with ultrapure water As a control solution, a solution containing about 2μg per 1ml. Under the chromatographic conditions described in Example 2, 25 μl of the reference solution was accurately measured and injected into the ion chromatograph to adjust the sensitivity of the instrument so that the chromatographic peak height of the main component was 10%-20% of the full scale. Then accurately measure 25μl each of the test solution and the reference solution, inject the samples separately, inject into the ion chromatograph, record the chr...
PUM
Abstract
Description
Claims
Application Information
- R&D Engineer
- R&D Manager
- IP Professional
- Industry Leading Data Capabilities
- Powerful AI technology
- Patent DNA Extraction
Browse by: Latest US Patents, China's latest patents, Technical Efficacy Thesaurus, Application Domain, Technology Topic, Popular Technical Reports.
© 2024 PatSnap. All rights reserved.Legal|Privacy policy|Modern Slavery Act Transparency Statement|Sitemap|About US| Contact US: help@patsnap.com