Optimization of electrostatically dosed dry powder inhaler
A fixed-dose and inhaler technology, applied in the direction of inhalers, respirators, therapeutic insufflators, etc., can solve the problems that the precise dosage is not well developed, and the device does not reach a high enough degree of decondensation.
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
preparation example Construction
[0071] Deagglomeration #2 is measured using the prepared DPI of step 140 .
[0072] Deagglomeration was then calculated using the electropowder particle size specification as input material, and HPLC analysis of the particle size distribution was performed as output after standard suction of the powder from the device components. Then compare the amount of powder less than 3 μm and 5 μm in the original electro-powder, and calculate the deagglomeration of the electro-dose as the percentage of de-agglomerated electro-dose at 3 μm, DD3 μm, 5 μm, and DD5 μm.
[0073] Figure 17 and 18 Calculations of deagglomeration at 3 μm and 5 μm are shown, respectively, and in the graphical representation the areas of the initial and final distributions are marked under the particle size distribution curve, respectively. The curve drawn with circles represents the initial electropowder size distribution, while the curve drawn with squares represents the final size distribution from the mouth...
example 1
[0089] Two different settings for the DPI intended in step 100 were prepared according to the following specifications, which were suitable for electrodosing in step 150 with 100 μg of albuterol (TBS) for local pulmonary delivery.
[0090] see Figure 6 , to determine the specification of the DPI by preparing and then analyzing the electrical dose. Figure 6 Indicates the decondensation of the electrical dose of TBS as a function of power in watts.
[0091] A typical set point for regional lung delivery is an inspiratory flow rate of 60 L / min, thus, from Figure 6 It is possible to calculate the pressure drop across the DPI that will achieve decondensation within the set specifications of the EDPI of the TBS.
[0092] By adjusting different sizes inside the DPI, it is possible to achieve the above set values. Figure 13 Represents a pictorial example of a mouthpiece with an electrical dose just below the opening. In this design it is possible, for example, to vary the dist...
example 2
[0172] Two different settings for the DPI planned in step 100 were prepared according to the following specifications, which were suitable for electrodosing in step 150 with 800 μg insulin (INS) for local pulmonary delivery.
[0173] see Figure 6 , to determine the specification of the DPI by preparing and then analyzing the electrical dose. Figure 6 Indicates the decondensation of the electrical dose of the INS as a function of power in watts.
[0174] A typical setpoint for deep lung delivery is an inspiratory flow rate of 40 L / min, thus, from Figure 6 It is possible to calculate the pressure drop across the DPI that will achieve decondensation within the set specifications of the EDPI of the INS.
[0175] DPI A
[0176] Suction flow rate 40 l / min
[0177] Suction pressure 1.5 kPa
[0178] Activation pressure 1.0 kPa
[0179] Dose delivery time t s 0.5 s
[0180] Dose delivery time T s 2.0 s
[0181] Activation time DPI T-t 3.0 s
[0182] Closin...
PUM
| Property | Measurement | Unit |
|---|---|---|
| Surface resistance | aaaaa | aaaaa |
Abstract
Description
Claims
Application Information
Login to View More 