PET calibrations with varying coincidence windows
Patent Information
- Authority / Receiving Office
- CN · China
- Current Assignee / Owner
- KONINKLIJKE PHILIPS ELECTRONICS NV
- Publication Date
- 2013-07-24
Smart Images
Figure 1 Figure 2 Figure 3
Abstract
Description
technical field
[0001] This application is particularly applicable to positron emission tomography (PET) systems, and in particular relates to PET scanner calibration. However, it will be appreciated that the described techniques may also find application in other medical imaging device calibration systems, other calibration scenarios, or other scanner calibration techniques. Background technique
[0002] As described in US Patent No. 7718954, a typical PET scanner performs a normalization calibration to correct for non-uniform 3D detector response, which typically takes about 6 hours. In addition, this scanner is calibrated with Standard Uptake Values ​​(SUV), which provides the conversion of counts in images to radioactive concentrations. The calibration used a decaying F-18 source and could have taken 11 hours or more to complete.
[0003] Any event measured by the PET scanner and occurring within a concurrent time window (e.g., 6ns), even two random single events, was ...