A method for determining the concentration of aniline in aromatic amines in workplaces
An aromatic amine, workplace technology, applied in the field of chemical compound detection, can solve the problems of increased column damage, column damage, high price, etc., to achieve the effect of ensuring accuracy, reducing usage, and ensuring precision and accuracy
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Examples
Embodiment 1
[0018] Adopt gas chromatography to measure the detection method of cyclohexylamine:
[0019] 1. Instrument
[0020] Silicone tube, solvent desorption type, filled with 200mg / 100mg silica gel.
[0021] Activated carbon tube, solvent desorption type, filled with 100mg / 50mg activated carbon.
[0022] Air sampler, flow rate 0~500ml / min.
[0023] Solvent analysis bottle, 5ml
[0024] Microsyringe, 10ml
[0025] Gas Chromatograph, Hydrogen Flame Ionization Detector
[0026] 2. Instrument testing conditions:
[0027] HP-INWWAX capillary column;
[0028] The column temperature is 180 degrees;
[0029] The vaporization chamber temperature is 220 degrees;
[0030] The temperature of the detection room is 240 degrees;
[0031] The carrier gas flow rate is 5ml / min.
[0032] 3. Reagents:
[0033] Desorption solution: absolute ethanol
[0034] 4. Sample collection:
[0035] Short-term sampling: At the sampling point, open both ends of the silicone tube and collect air samples f...
Embodiment 2
[0060] Precision measurement experiment of the present invention:
[0061] Draw four samples of 30 μl into four 1ml analysis bottles, respectively labeled 183, 184, 185 and 186, and carry out the determination of the sample concentration by the sample determination method of Example 1, the results are as follows:
[0062] Sample number A (absorbance) C (concentration μg / ml) Aniline 183 297.356 378.482 Aniline 184 297.391 378.528 Aniline 185 297.463 378.619 Aniline 186 297.427 378.573
[0063] After calculation, the average concentration of aniline is 378.551μg / ml, and the results of the sample show that the precision test meets the requirements of the standard.
[0064] After setting the appropriate measurement conditions in this example, the detection of other compounds in aromatic compounds can also be carried out at the same time, which solves the problem of frequent column replacement and easy damage to the column and time-consu...
Embodiment 3
[0066] Spike recovery experiment:
[0067] Pipette 30μl of 12.58mg / ml concentrated standard and 30μl of aniline sample into 1ml analysis bottle respectively, and the recoveries of each standard addition are shown in the table below:
[0068] Sample number A C Aniline 1 305.337 388.638 Aniline 2 308.180 392.292 Aniline 3 308.263 392.358 Aniline 4 305.295 388.582
[0069] Calculated: the recovery rate of the addition of sample 1 is 99.97%;
[0070] The standard recovery rate of sample 2 was 100.93%;
[0071] The standard recovery rate of sample 3 was 100.95%;
[0072] Spike recovery for sample 4 was 99.95%
PUM
Login to View More Abstract
Description
Claims
Application Information
Login to View More