Embodiment 1
[0141] 1. Test process
[0142] S1. Set a series of different Ca / Zr molar ratios, mix zirconium dioxide and calcium carbonate according to the set ratios, and obtain a series of standard products with different calcium-zirconium molar ratios.
[0143] The setting values are as follows: 0.9650, 0.9699, 0.9752, 0.9802, 0.9851, 0.9902, 0.9950, 0.9998.
[0144] S2. Treat the standard product at 150°C for 30 minutes to obtain the pre-treated standard product.
[0145] Weigh 0.6000g±0.0001g pretreatment standard and 6.0000g±0.0001g flux (lithium tetraborate 67%, lithium metaborate 33%), put them into a platinum crucible, and stir evenly. Then, 0.5±0.05g of mold release agent (anhydrous lithium bromide solution, concentration 30.5g / 50mL) was added dropwise. After that, put it into a high-frequency fusion machine for sample preparation, and the conditions are: the melting temperature is 1170°C, the pre-stand time is 240s, and the swing time is 600s. Get a standard fuse.
[0146]...
Embodiment 2
[0174] 1. Test process
[0175] S1. Set a series of different Ca / Zr molar ratios, mix zirconium dioxide and calcium carbonate according to the set ratios, and obtain a series of standard products with different Ca / Zr molar ratios.
[0176] The setting values are as follows: 0.9650, 0.9699, 0.9752, 0.9802, 0.9851, 0.9902, 0.9950, 0.999.
[0177] S2. Treat the standard product at 150°C for 30 minutes to obtain the pre-treated standard product.
[0178] Weigh 0.6000g±0.0001g pretreatment standard and 6.0000g±0.0001g flux (lithium tetraborate 67%, lithium metaborate 33%), put them into a platinum crucible, and stir evenly. Then, 0.5±0.05g of mold release agent (anhydrous lithium bromide solution, concentration 30.5g / 50mL) was added dropwise. After that, put it into a high-frequency fusion machine for sample preparation, and the conditions are: the melting temperature is 1170°C, the pre-stand time is 240s, and the swing time is 600s. Get a standard fuse.
[0179] S3. Construc...