Embodiment 1
[0029] The determination of 1,3-propane sultone content in the dyestuff sample of embodiment 1
[0030] (1) Analysis steps
[0031] Take 50g dye sample and mix well. Weigh 1.03g and 1.01g samples respectively into 50mL glass centrifuge tubes, add 20mL n-hexane, vortex mix, room temperature ultrasonic extraction for 30min, 8000r / min high-speed centrifugation for 10min, take the supernatant, repeat extraction with 10mL n-hexane to combine the upper layer For the supernatant, use n-hexane to dilute the supernatant to 50mL as the solution to be tested, centrifuge at 8000r / min for 10min at a high speed, measure 2mL at the middle scale of the solution to be tested with a pipette, filter through a 0.22μm nylon filter membrane, and filter the filtrate through Gas chromatography-mass spectrometry analysis.
[0032] (2) Chromatography and mass spectrometry parameters
[0033] Gas chromatographic conditions, adopt chromatographic column DB-5MS, 30m×0.25mm×0.25μm, take 60℃ as the initi...
Embodiment 2
[0039] Embodiment 2 dyestuff sample 1, the standard substance of 3-propane sultone adds and measures
[0040] (1) Analysis steps
[0041] Take the 1,3-propane sultone negative sample in Example 1. Weigh six samples of about 1g (accurate to three decimal places) into a 50mL glass centrifuge tube, add 0.1mL each of 1000μg / mL 1,3-propane sultone standard stock solution, after standing still for a period of time, add 20mL n-hexane Alkanes, 8000r / min high-speed centrifugation for 10min, take the supernatant, repeat extraction with 10mL n-hexane to combine the supernatant, use n-hexane to dilute the supernatant to 50mL as the solution to be tested, 8000r / min high-speed centrifugation for 10min, pipette Measure 2mL of the tube at the middle scale of the solution to be tested, filter it through a 0.22μm nylon filter membrane, and analyze the filtrate by gas chromatography-mass spectrometry.
[0042] (2) Instrument conditions
[0043]Gas chromatographic conditions, adopt chromatogra...