Method for selecting duodenal fluid sample for detecting pancreatic disease marker, and method for detecting pancreatic disease marker
A duodenal fluid and determination method technology, which is applied in the field of duodenal fluid samples, can solve problems such as small difference and poor detection accuracy, and achieve the effects of reducing missed detection, suppressing unnecessary inspection, and improving inspection performance
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment
[0113] Next, the present invention will be described in more detail by showing examples and the like, but the present invention is not limited to the following examples.
reference example 1
[0115] Duodenal fluid samples were collected endoscopically from 49 pancreatic cancer cases, and the CEA concentration in the samples was measured. The CEA concentration was measured by the ELISA method using a commercially available kit (manufactured by IBL). Based on the measured CEA concentration, the critical value was set at 126ng / mL to judge negative and positive.
[0116] The sensitivity and the ratio of the number of positive samples to the total number of samples obtained from the test results are shown in Table 1 for each T classification indicating the degree of local development of the pancreas of the main lesion. Tis refers to non-invasive cancer; T1 refers to tumor diameter less than 2 cm and confined to the pancreas; T2 refers to tumor diameter exceeding 2 cm and confined to the pancreas; T3 refers to cancer invasion involving the bile duct in the pancreas, duodenum, and surrounding tissues of the pancreas Any one; T4 refers to any one in which the cancer infil...
Embodiment 1
[0121] Absorption spectra were measured for duodenal fluid samples collected endoscopically from 49 cases of pancreatic cancer and 7 cases of benign pancreatic diseases. The absorption spectra of all samples to be tested are shown in Figure 4 . Will Figure 4 The magnified diagram near the absorption peak (300-560nm) in Figure 5 . It can be seen that the duodenal fluid sample has an absorption peak wavelength region at 350 to 540 nm, especially at 400 to 460 nm.
[0122] Furthermore, the density of the color of each duodenal fluid sample (specimen) was evaluated. Specifically, using a spectrophotometer, the absorption spectrum (A 10 ). Next, based on the following formula, the obtained absorption spectrum was converted into a transmission spectrum (T 10 ), and further take the optical path length 5mm (T 5 ) for conversion.
[0123] T 10 = 10 - A ...
PUM
| Property | Measurement | Unit |
|---|---|---|
| diameter | aaaaa | aaaaa |
| Sensitivity | aaaaa | aaaaa |
| Sensitivity | aaaaa | aaaaa |
Abstract
Description
Claims
Application Information
Login to View More 