Preparation and detection method of hepatobiliary acid determination kit
A kit and reagent technology, applied in the field of pharmaceutical reagents, can solve the problems of high cost, long measurement time, errors, etc.
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0032] Example 1 Preparation of hepatic bile acid detection kit
[0033] The hepatic bile acid detection kit prepared according to the present invention comprises independently packaged reagent R1 and reagent R2, and the reagent R1 comprises components of the following concentrations:
[0034] Glycine buffer: 100mmol / L;
[0035] Sodium chloride: 9g / L;
[0036] Triton X-100: 0.5g / L;
[0037] PEG20000: 20g / L
[0038] PC300: 1g / L;
[0039] Described reagent R2 comprises the component of following concentration:
[0040] Glycine buffer: 100mmol / L;
[0041] Antibody latex particles: 35mg / L;
[0042] Sucrose: 5g / L;
[0043] PC300: 1g / L;
[0044] Preferably, the preparation steps of the antibody latex particles in the reagent R2 are:
[0045] Preparation: prepare 220nm polystyrene latex particles, and dilute the latex particles with 5mM / L 2-(N-morpholine)ethanesulfonic acid at pH 7.0 to a final concentration of 1.5% w / v to obtain a latex particle solution; Goat polyclonal a...
Embodiment 2
[0049] Embodiment 2 Analysis of the accuracy of the kit of the present invention
[0050] Test kit: the kit of Example 1, the hepatocholic acid assay kit of the comparative example (produced by Beijing Jiuqiang Biotechnology Co., Ltd.)
[0051] Test instrument: Hitachi 7100 automatic biochemical analyzer
[0052] The specific operation is to set parameters: test method: two-point end point method; main wavelength: 600nm; photometry point: 18-34; sample size: 6; reagent volume R1: R2=150: 50; calibration method: spline
[0053] Setting of the standard curve: Put the standard substances of known concentration, namely 5, 10, 20, 40, 80 mg / L in the calibration plate, set the position and concentration of the calibration substance, and obtain the following data:
[0054] Table 1 (Example 1)
[0055]
[0056] Table 2 (comparative example)
[0057]
[0058] 40 cases of clinical samples with different concentrations were tested in parallel with the control, and the following ...
Embodiment 3
[0064] Embodiment 3 Precision analysis of the kit of the present invention
[0065] Test kit: the kit of Example 1, the hepatocholic acid assay kit of the comparative example (produced by Beijing Jiuqiang Biotechnology Co., Ltd.)
[0066] Test instrument: Hitachi 7100 automatic biochemical analyzer
[0067] The specific operation is to set parameters: test method: two-point end point method; main wavelength: 600nm; photometry point: 18-34; sample size: 6; reagent volume R1: R2=150: 50; calibration method: spline
[0068] After calibration, the same serum sample was measured for 10 times, and the average value and SD value were obtained from the concentration results of 10 times of detection, and the CV value was calculated. The results are shown in Table 5:
[0069]
[0070] As can be seen from the data in Table 5, the repeatability of the measurement results of the same sample in Example 1 is almost the same as that of the comparative example, indicating that this reagen...
PUM
Login to View More Abstract
Description
Claims
Application Information
Login to View More 


