Vacuum sampling device for gas chromatograph
A gas chromatograph and sampling device technology, applied in the field of vacuum sampling devices for gas chromatographs, can solve problems such as interference analysis results, detector response overload, analysis result errors, etc., to expand the response linear range and improve analysis accuracy , the effect of high analysis accuracy
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0027] Take a helium sample gas tank, install it on the position of the sample container 2, and measure the volume concentration of each impurity component in the helium sample according to the following steps.
[0028] 1a. Turn on the power switch of the housing 19, open the pressure sensor valve 11 and the vacuum gauge valve 14, and obtain the air pressure value, vacuum degree and temperature value from the pressure display instrument 9, the vacuum display instrument 16 and the temperature display instrument 7, and the temperature value is 19.9°C;
[0029] 1b. Close calibration gas valve 5, vacuum valve 17 and vent valve 13;
[0030] 1c. Start the vacuum pump 18, open the vacuum valve 17 and the sample gas valve 6 in turn, and start vacuuming;
[0031] 1d. When the reading of the vacuum display instrument 16 is lower than 5Pa, close the vacuum valve 17, open the switch of the sample container 2 until the required sample gas enters the trachea, close the switch of the sample...
Embodiment 2
[0036] Take a standard gas cylinder whose balance gas is He and whose components are 5ppmH2, 6.1ppmO2, and 7.5ppmN2, install it on the position of standard gas cylinder 1, and measure the peak area of each component chromatographic peak in the standard gas according to the following steps .
[0037] 2a. Turn on the power switch of the housing 19, open the pressure sensor valve 11 and the vacuum gauge valve 14, and obtain the air pressure value, vacuum degree and temperature value from the pressure display instrument 9, the vacuum display instrument 16 and the temperature display instrument 7, and the temperature value is 19.9°C;
[0038] 2b. Close the sample gas valve 6 and the vacuum valve 17, open the switch of the standard gas cylinder 1, the standard gas valve 5 and the vent valve 13, adjust the pressure reducing valve 3, and use the standard gas to purge for 3-5 minutes;
[0039] 2c. Close the calibration gas valve 5 and the vent valve 13, start the vacuum pump 18, ope...
Embodiment 3
[0045]The implementation of this embodiment is basically the same as that of Embodiment 2, the main difference being that, considering the interference of oxygen and nitrogen in the atmosphere during the measurement process, the oxygen and nitrogen content in high-purity helium with a purity higher than 99.999% was measured. Take a high-purity helium cylinder with a purity higher than 99.999%, install it on the position of the standard gas cylinder 1, measure and analyze the oxygen and nitrogen content, and obtain the background oxygen and nitrogen volume concentration in the measurement step as O2: 0.06ppm, N2 : 0.25ppm.
PUM
| Property | Measurement | Unit |
|---|---|---|
| diameter | aaaaa | aaaaa |
Abstract
Description
Claims
Application Information
Login to View More 
