A kind of midkine midkine biological analysis method and detection kit
A technology for detection kits and midkine, applied in biological testing, biological material analysis, analytical materials, etc., can solve the problems of midkine content differences, heparin capture, midkine detection can not meet the requirements, etc., to broaden the detection range and improve reactivity , the effect of eliminating interference
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0126] Use Assay Buffer, 10% serum (that is, the volume ratio of serum to Assay Buffer is 1:9), and serum to prepare a standard curve, and compare the differences in detection signals. See the table below for details:
[0127]
[0128] It can be seen that the signal of the standard prepared in Buffer is much higher than that in serum, and the signal decreases with the increase of serum concentration. This once again confirms that the measurement results of the serum samples obtained by directly preparing the standard in the buffer of the current kit will be low. Therefore, the final standard curve of the detection method and the kit of the present invention are consistent with the quality control and the sample to be tested, and the blood sample cannot be directly prepared with buffer as in the conventional detection method.
[0129] figure 1 It is the standard curve of the midkine Midkine biological analysis method of the present invention, wherein Std1 is the standard c...
Embodiment 2
[0131] Example 2: Setting up a control experiment without acidification to determine the effect of acidification on the removal of heparin. See the table below for details:
[0132]
[0133] We use Assay Buffer and 10% serum to prepare the standard song, because the signal in the Assay Buffer is very high, it is used as a positive control in this experiment, and the standard song prepared with 10% serum is acidified with acetic acid and hydrochloric acid respectively, according to the experiment According to the result data, it can be found that the signal of Assay can be obviously promoted when using 300mM acetic acid for acidification treatment, and the highest signal is equal to that of Buffer, indicating that this can eliminate the interference of heparin, but the acidification effect of hydrochloric acid is not very ideal, so the method protected in the invention Acetic acid is used as the acidifying solution.
Embodiment 3
[0134] Embodiment 3: Based on the above-mentioned groping test conditions, each concentration point of the standard curve was prepared with serum, and each concentration point of QC was also prepared with serum, so that the standard curve and QC were prepared for the recovery test. See the table below for details:
[0135]
[0136] (NA does not participate in the fitting in this experiment)
[0137] The following two groups of QC are prepared from different individual sera. See the table below for details:
[0138]
[0139] From this group of experimental data, it can be seen that the signal is normal after acidification treatment, and the QC recovery rate of one serum preparation is very good within 70-125%, but the other fluctuates greatly, indicating that the method also needs to deal with different matrices The problem of fluctuations, the problem of matrix interference.
PUM
Login to View More Abstract
Description
Claims
Application Information
Login to View More 


