Kit for detecting five series viruses of infant hand-foot-and-mouth disease and application of kit
A virus detection and kit technology, which is used in the determination/inspection of microorganisms, DNA/RNA fragments, recombinant DNA technology, etc., can solve the problems of high reagent price, long detection time, and inability to achieve instant inspection, and achieve good application prospects. , high sensitivity and specificity
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0107] Example 1 Repeatability Detection
[0108] The positive control of each virus diluted 10 times was used as the sample to be tested for repeated detection, and the detection was repeated 3 times.
[0109] 1. Nucleic acid extraction (used with Hs480 heating centrifuge)
[0110] 1) Take 50 μl of the sample and add 5 μl FastLyseL4, shake and mix;
[0111] 2) Place the mixed sample in the HS480 heating centrifuge machine, heat at 95°C for 2min, and centrifuge at 5000rpm for 2min;
[0112] 3) After centrifugation, take the supernatant as the sample RNA, numbered;
[0113] 4) The positive control, negative control and samples to be tested are processed synchronously.
[0114] 2. Preparation of PCR reagents
[0115] 1) Prepare the PCR reaction solution: pre-mix the ultra-fast PCR buffer1 (68 μl) and the ultra-fast PCR enzyme system (17 μl) (2 more than the actual number of reaction tubes, to avoid loss during sample addition, and to avoid air bubbles during mixing);
[011...
Embodiment 2
[0134] Embodiment 2 specific detection
[0135] Each virus positive control was used as the sample to be tested for specific detection, and the detection method was the same as in Example 1.
[0136] Result: Each virus positive control sample was only positive under the corresponding probe, and the rest were negative, with good specificity. The results were as follows: Figure 1-Figure 5 shown.
Embodiment 3
[0137] Example 3 Sensitivity Detection
[0138] Each virus positive control in the kit (concentration is 1 × 10 5 copies / ml) was serially diluted to 1×10 4 copies / ml, 1×10 3 copies / ml, 1×10 2 copies / ml and 1×10 1 Copies / ml, after preparation, use it as the sample to be tested for sensitivity testing with the corresponding kit.
[0139] Concrete implementation steps are the same as embodiment 1
[0140] Results: The detection limit of CA6 is not higher than 1×10 2 copies / ml, the detection limit of CA10, CA16, EV71 and EVU is not higher than 1×10 1 copies / ml, the result is as Image 6 shown.
PUM
Login to View More Abstract
Description
Claims
Application Information
Login to View More 


