HPLC analysis method of bortezomib related substances
An analysis method and technology related to substances, applied in the direction of analysis of materials, material separation, measurement devices, etc., can solve problems such as large errors, affecting the quality of anti-tumor drug bortezomib, and difficulty in gasification
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0037] 1) Chromatographic conditions:
[0038] Instrument: Agilent1260HPLC
[0039] Mobile phase: 10mmol / L of K 2 HPO 4 Buffer (adjust pH to 5.0 with phosphoric acid)-acetonitrile (volume ratio 55:45)
[0040] Column: OX-3 (4.6×150mm, 3μm)
[0041] Detection wavelength: 235nm
[0042] Column temperature: 25°C
[0043] Flow rate: 0.4ml / min
[0044] Injection volume: 5ul
[0045] 2) Sample preparation:
[0046] Take about 50 mg of (1R)-(S)-pinanediol-1-ammonium trifluoroacetate-3-methylbutane-1-boronate (SCR-3081), weigh it accurately, and place it in a 25ml measuring bottle , add 20 μl of benzoyl chloride and shake well, then add 50 μl of DBU (1,8-diazabicyclo[5,4,0]undec-7-ene), add 15ml of acetonitrile to dissolve by ultrasound, and react in a water bath at 50°C for 30 minutes. Allow to cool to room temperature, dilute to volume with methanol, shake well, filter, and take the filtrate as the test solution.
[0047] Accurately weigh 5 mg each of SCR-5220 and SCR-52...
Embodiment 2
[0051] 1) Chromatographic conditions:
[0052] Instrument: Agilent1260HPLC
[0053] Mobile phase: phosphate buffered saline (pH 3.0)-acetonitrile (volume ratio 65:35)
[0054] Column: OX-3 (4.6×150mm, 3μm)
[0055] Detection wavelength: 235nm
[0056] Column temperature: 20°C
[0057] Flow rate: 1.0ml / min
[0058] Injection volume: 5ul
[0059] 2) Sample preparation:
[0060]Take about 50 mg of (1R)-(S)-pinanediol-1-ammonium trifluoroacetate-3-methylbutane-1-boronate (SCR-3081), weigh it accurately, and place it in a 25ml measuring bottle , add 20, add benzoyl chloride and shake well after adding 50 acid chloride (1,8 diazabicyclo[5,4,0]undec-7-ene), add 15ml of acetonitrile to dissolve by ultrasound, at 60 React in a water bath at ℃ for 30 minutes, let cool to room temperature, dilute to the mark with methanol, shake well, filter, take the subsequent filtrate as the test solution.
[0061] Accurately weigh 5 mg each of SCR-5220 and SCR-5221 reference substances, put...
Embodiment 3
[0065] 1) Chromatographic conditions:
[0066] Instrument: Agilent1260HPLC
[0067] Mobile Phase: Phosphate Buffered Saline (pH 6.5)-Acetonitrile (40:60)
[0068] Column: OX-3 (4.6×150mm, 3μm)
[0069] Detection wavelength: 235nm
[0070] Column temperature: 30°C
[0071] Flow rate: 0.2ml / min
[0072] Injection volume: 5ul
[0073] 2) Sample preparation:
[0074] Take about 50 mg of (1R)-(S)-pinanediol-1-ammonium trifluoroacetate-3-methylbutane-1-boronate (SCR-3081), weigh it accurately, and place it in a 25ml measuring bottle , add 20 μl of benzoyl chloride, shake well, add 50 μl of DBU (1,8-diazabicyclo[5,4,0]undec-7-ene), add 15ml of acetonitrile to dissolve it by ultrasound, and react in a water bath at 45°C for 30 minutes. Allow to cool to room temperature, dilute to volume with methanol, shake well, filter, and take the filtrate as the test solution.
[0075] Accurately weigh 5 mg each of SCR-5220 and SCR-5221 reference substances, put them in a 25ml measuring ...
PUM
| Property | Measurement | Unit |
|---|---|---|
| volume | aaaaa | aaaaa |
| volume | aaaaa | aaaaa |
| wavelength | aaaaa | aaaaa |
Abstract
Description
Claims
Application Information
Login to View More 