A technology of optical purity and pasireotide, which is applied in the field of optical purity analysis of pasireotide starting materials, can solve the problems of ineffective control of the optical purity of amino acids, no chromogenic groups in molecules, large method errors, etc. Fast, high sensitivity, fast and accurate analysis, convenient effect
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0120] The optical purity test method of N(e)-tert-butoxycarbonyllysine:
[0121] Chromatographic column: ChiralPak ZWIX(+), 4×150 nm, 3 μm.
[0122] Mobile phase: A: methanol; B: acetonitrile; methanol: acetonitrile (40:60) + 50mM formic acid + 25mM diethylamine;
[0123] Flow rate: 0.4mL / min;
[0124] Column temperature: 25°C;
[0125] Elution method: isocratic elution;
[0126] Detection wavelength: 220nm;
[0127] Injection volume: 10 μL.
[0128] Preparation of the control sample solution: Accurately weigh appropriate amounts of (S)-type N(e)-tert-butoxycarbonyllysine and (R)-type N(e)-tert-butoxycarbonyllysine standard substances respectively, add ACN: H 2 O(1:1)+1%TFA dissolved and diluted to make about 10mg of S-type N(e)-tert-butoxycarbonyllysine and R-type N(e)-tert-butoxycarbonyllysine per 1mL Amino acid was passed through a 0.45 μm microporous membrane as a reference solution.
[0129] Configuration of the test sample solution: Accurately weigh an appropria...
Embodiment 2
[0134] The optical purity test method of N(e)-tert-butoxycarbonyllysine:
[0135] Chromatographic column: ChiralPak ZWIX(+), 250×4 nm, 3 μm.
[0136] Mobile phase: A: methanol; B: acetonitrile; methanol: acetonitrile (40:60) + 50mM acetic acid + 25mM triethylamine solution;
[0137] Flow rate: 0.6mL / min;
[0138] Column temperature: 30°C;
[0139] Elution method: isocratic elution;
[0140] Detection wavelength: 220nm;
[0141] Injection volume: 10 μL.
[0142] Preparation of the control sample solution: Accurately weigh appropriate amounts of (S)-type N(e)-tert-butoxycarbonyllysine and (R)-type N(e)-tert-butoxycarbonyllysine standard substances respectively, add ACN: H 2 O(1:1)+1%TFA dissolved and diluted to make about 20mg of (S)-type N(e)-tert-butoxycarbonyllysine and R-type N(e)-tert-butoxylysine per 1mL Carbonyl lysine was passed through a 0.45 μm microporous membrane as a reference solution.
[0143] Configuration of the test sample solution: Accurately weigh an ...
Embodiment 3
[0148] The optical purity test method of N(e)-tert-butoxycarbonyllysine:
[0149] Chromatographic column: ChiralPak ZWIX(+), 200×4.6 nm, 3 μm.
[0150] Mobile phase: A: methanol; B: acetonitrile; methanol: acetonitrile (40:60) + 50mM propionic acid + 25mM ammonia solution;
[0151] Flow rate: 0.2mL / min;
[0152] Column temperature: 20°C;
[0153] Elution method: isocratic elution;
[0154] Detection wavelength: 220nm;
[0155] Injection volume: 10 μL.
[0156] Preparation of the control sample solution: Accurately weigh appropriate amounts of (S)-type N(e)-tert-butoxycarbonyllysine and (R)-type N(e)-tert-butoxycarbonyllysine standard substances respectively, add ACN: H 2 O(1:1)+1%TFA dissolved and diluted to make about 15mg (S)-type N(e)-tert-butoxycarbonyllysine and R-type N(e)-tert-butoxylysine per 1mL Carbonyl lysine was passed through a 0.45 μm microporous membrane as a reference solution.
[0157] Configuration of the test sample solution: Accurately weigh an appr...