Method for monitoring permeation behavior of photoinitiator 907 in plastic package
A photoinitiator, -907 technology, applied in measuring devices, analytical materials, material excitation analysis, etc., can solve problems such as human health hazards, food pollution, and weak signals
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0038] In this embodiment, seven commonly used photoinitiators are selected, namely ITX, TPO, EDB, EHA, Irgacure 907, MK and CBP, and their Raman spectra are collected.
[0039] Pipette 2 μL of seven photoinitiator standard solutions (ITX, TPO, EDB, EHA, Irgacure 907, MK, and CBP) with a concentration of 10 mg / L and transfer them to the parafilm, blank acetonitrile as a control, and then add 2 μL of AuNPs (50nm, 250 mg / L), and mixed with a pipette gun for 20 s, then 2 μL of the mixed solution was dropped onto a gold plate, and dried at room temperature in the dark for Raman spectrum measurement.
[0040] The standard solutions (10 mg / L) of these seven photoinitiators were dropped onto the gold plate for SERS detection. The SERS spectra of seven photoinitiators are shown in figure 1 , compared with the control, the photoinitiators with Raman signal include Irgacure 907, TPO, MK and ITX, while EDB, EHA and CBP have no Raman signal. The characteristic peak of Irgacure 907 is at...
Embodiment 2
[0049] Based on the theoretical results of Example 1, this example provides a method for the penetration behavior of photoinitiator-907 in PE, and the specific operations are as follows:
[0050] Prepare 200mg / L Irgacure 907 acetone solution, take 2 μL Irgacure 907 acetone solution and drop it on the front of PE, and dry it in a closed petri dish wrapped with tinfoil at room temperature. Take it out at 6h and 12h.
[0051] Prepare a 250mg / LAuNPs solution, drop 50nm on the same position of PE plastic, and scan at a depth of 0-40μm; add AuNPs at the position where the photoinitiator is exposed on the reverse side of the plastic at the same time, and scan at a depth of 40-80μm. The parameters of the Raman instrument are set as laser emission power of 0.5mW, acquisition time of 0.5s, scanning range of 5×5μm, 9 plane scanning points, and a step size of 20μm.
[0052] Scan results see Figure 9 ,Depend on Figure 9 It can be seen that under this condition, the penetration of phot...
Embodiment 3
[0054] This embodiment provides a method for the penetration behavior of photoinitiator-907 in PET, and the specific operations are as follows:
[0055] Dissolve 200 mg / L photoinitiator-907 (Irgacure 907) solution in acetonitrile, then take 2 μL and drop it on the front of PET, and dry it in an airtight Petri dish wrapped with tinfoil at room temperature, with different exposure times (1d, 5d, 10d, 15d, 20d) were taken out, and AuNPs (50nm, 250mg / L) was dropped to the same position on the front of the plastic, and the scanning depth was 0-40μm; at the position where the photoinitiator-907 was exposed on the back of the plastic At the same time, AuNPs were added dropwise, and the scanning depth was 40-80 μm. The parameters of the Raman instrument were set to 0.5mW power, 0.5s acquisition time, scanning range 5×5μm, 9 plane scanning points, and a step diameter of 20μm.
[0056] Scan results see Figure 10 ,Depend on Figure 10 It can be seen that under this condition, the pen...
PUM
Property | Measurement | Unit |
---|---|---|
concentration | aaaaa | aaaaa |
Abstract
Description
Claims
Application Information

- R&D
- Intellectual Property
- Life Sciences
- Materials
- Tech Scout
- Unparalleled Data Quality
- Higher Quality Content
- 60% Fewer Hallucinations
Browse by: Latest US Patents, China's latest patents, Technical Efficacy Thesaurus, Application Domain, Technology Topic, Popular Technical Reports.
© 2025 PatSnap. All rights reserved.Legal|Privacy policy|Modern Slavery Act Transparency Statement|Sitemap|About US| Contact US: help@patsnap.com